The HydraFacial® MD handpiece tubing chambers must be cleaned and disinfected using the cleaning brush every week to prevent any bacterial contamination.
HydraFacial® Handpiece
The HydraFacial® handpiece tubing chambers must be cleaned and disinfected using the cleaning brush every week to prevent any bacterial contamination.
Note: Make sure the flow control is fully open (+)
- Insert the cleaning brush into the vacuum (smaller) opening.
- Insert the cleaning brush into the vacuum (smaller opening) on the face of the hand piece. Push it in gently all the way, while rotating in either direction. Repeat with the opening on the back with the blue dot. A few drops of disinfectant may be dropped onto the “face” of the hand piece (held in an upright position) allowing the disinfec
tant to run into the handpiece opening before inserting the cleaning brush. Note: 70% isopropyl alcohol is not suitable as a disinfectant.
- Wash the brush bristles in soapy water between each handpiece
opening insertion to clean off debris.
- Using a compressed air canister with the straw attached, blow air
first into the vacuum (smaller opening) on the face of the handpiece, and then into the opening in the gray cartridge on the back of the hand piece.
- When the brush and air cleaning are completed, make sure all the blue toggle switches are in the OFF position. Turn the gray toggle
switch “SYSTEM RINSE” up to the ON position and the vacuum setting on “26”. Press on the cleaning cap and flush through the handpiece and tubing for I minute.